Portable Turbidity Meter TU200P/TU1000P
Bench Top Turbidity Meter TU200/TU1000/TU4000E
TU200P Portable Turbidity Meter
GENERAL FEATURE
² Uses Nephelometric principle with 90° detection.
² Meets ISO 7027 and offers consistent light source for higher accuracy.
² Selectable turbidity units, including the NTU, FNU and EBC.
² Settable parameters, including user ID, number of calibration points, data and time, etc.
² Auto-power-off feature effectively extends the battery usage time.
² IP65 waterproof. The portable meter is suitable for fields measurements and out door measurements.
² 1-4 points calibration using Formazin standards or zero water.
² Autoranging from 0-200 NTUs.
PRODUCT SPECIFICATIONS
Model | TU200P | |
Measuring Unit | Range | (0~20.00) NTU, (20.0~200.0) NTU |
Resolution | 0.01NTU, 0.1NTU, 1NTU | |
Accuracy | ≤±8% | |
Repeatability | ≤1.0% | |
Fluctuation | ≤±1.0% FS/30min | |
General | Power Supply | AA Battery, AC Adapter, 200-240 VAC input,DC 5V output |
Dimension (mm)& Weight (kg) | 100×220×80,0.8 |
TU1000P Portable Turbidity Meter
GENERAL FEATURE
² Uses Nephelometric principle with 90° detection.
² Meets ISO 7027 and offers consistent light source for higher accuracy.
² Selectable turbidity units, including the NTU, FNU and EBC.
² Settable parameters, including user ID, number of calibration points, data and time, etc.
² Auto-power-off feature effectively extends the battery usage time.
² IP65 waterproof. The portable meter is suitable for fields measurements and out door measurements.
² 2 photodiodes to ensure high accuracy.
² 1-7 points calibration using Formazin standards or zero water.
² Autoranging from 0-1000 NTUs.
² Data capacity of up to 2000 sets (GLP-compliant).
² Data stored can be transferred into the computer by USB or RS-232 communication interface.
PRODUCT SPECIFICATIONS
Model | TU1000P | |
Measuring Unit | Range | (0~20.00) NTU, (20.0~200.0) NTU (200~1000) NTU |
Resolution | 0.01NTU, 0.1NTU, 1NTU | |
Accuracy | ≤±6% | |
Repeatability | ≤0.5% | |
Fluctuation | ≤±0.5% FS/30min | |
General | Power Supply | AA Battery, AC Adapter, 200-240 VAC input,DC 5V output |
Dimension (mm)& Weight (kg) | 100×220×80,0.8 |
TU200 Bench Top Turbidity Meter
GENERAL FEATURES
² Uses Nephelometric principle with 90° detection.
² Meets ISO 7027 and offers consistent light source for higher accuracy.
² Selectable turbidity units, including the NTU, FNU and EBC.
² Settable parameters, including user ID, number of calibration points, data and time, etc.
² 1-5 points calibration using Formazin standards or zero water.
² Autoranging from 0-200 NTUs.
PRODUCT SPECIFICATIONS
Model | TU200 | |
Measuring Unit | Range | (0~20.00) NTU, (20.0~200.0) NTU |
Accuracy | ≤±8% | |
Repeatability | ≤1.0% | |
Fluctuation | ≤±1.0% FS/30min | |
General | Power Supply | AC Adapter, 200-240 VAC input |
Dimension (mm)& Weight (kg) | 330×250×130, 2.5 |
TU1000 Bench Top Turbidity Meter
GENERAL FEATURES
² Uses Nephelometric principle with 90° detection.
² Meets ISO 7027 and offers consistent light source for higher accuracy.
² Selectable turbidity units, including the NTU, FNU and EBC.
² Settable parameters, including user ID, number of calibration points, data and time, etc.
² 2 photodiodes to ensure high accuracy.
² 1-7 points calibration using Formazin standards or zero water.
² Autoranging from 0-1000 NTUs.
² Data capacity of up to 2000 sets (GLP-compliant).
² Data stored can be transferred into the computer by USB or RS-232 communication interface.
PRODUCT SPECIFICATIONS
Model | TU1000 | |
Measuring Unit | Range | (0~20.00) NTU, (20.0~200.0) NTU (200~1000) NTU |
Accuracy | ≤±6% | |
Repeatability | ≤0.5% | |
Fluctuation | ≤±0.5%FS/30min | |
General | Power Supply | AC Adapter, 200-240 VAC input |
Dimension (mm)& Weight (kg) | 330×250×130, 2.5 |
TU4000E Bench Top Turbidity Meter
GENERAL FEATURES
² 7” colorful touchscreen.
² Tungsten lamp light source met USEPA 180.1 requirement.
² 3 photodiodes to ensure high accuracy.
² 1-8 points calibration using Formazin standards or zero water.
² Autoranging from 0-4000 NTUs.
² Data Storage capacity of 1000 sets (GLP-compliant).
² Data stored can be transferred into the computer by USB or RS-232 communication interface.
² Reset feature automatically resumes all settings back to factory default options.
PRODUCT SPECIFICATIONS
Model | TU4000E | |
Measuring Unit | Range | (0.000~20.00) NTU, (20.01~200.0) NTU, (200~2000) NTU, (2001~4000) NTU |
Accuracy | ≤±6% | |
Repeatability | ≤0.5% | |
Fluctuation | ≤±0.5% FS/30min | |
General | Power Supply | AC Adapter, 200-240 VAC input |
Dimension (mm)& Weight (kg) | 358×323×160, 8 |