TopPette Pipette/Lab pipette/low price pipette/digital pipette
ME-TopPette Pipette
Features:
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied
with individual test certificate
- The low part is available for autoclaving
Specifications:
Single Chanel | Volume Range | Increment | Test Volume(μl) | Accuracy error | Precision error | ||
Model | % | μl | % | μl | |||
ME-TopPette(1/0.1-2.5μl ) | 0.1-2.5μl | 0.05μl | 2.5 | 2.50% | 0.0625 | 2.00% | 0.05 |
1.25 | 3.00% | 0.0375 | 3.00% | 0.0375 | |||
0.25 | 12.00% | 0.03 | 6.00% | 0.015 | |||
ME-TopPette(1/0.5-10μl) | 0.5-10μl | 0.1μl | 10 | 1.00% | 0.1 | 0.80% | 0.08 |
5 | 1.50% | 0.075 | 1.50% | 0.075 | |||
1 | 2.50% | 0.025 | 1.50% | 0.015 | |||
ME-TopPette(1/2-20μl) | 2-20μl | 0.5μl | 20 | 0.90% | 0.18 | 0.40% | 0.08 |
10 | 1.20% | 0.12 | 1.00% | 0.1 | |||
2 | 3.00% | 0.06 | 2.00% | 0.04 | |||
ME-TopPette(1/5-50μl) | 5-50μl | 0.5μl | 50 | 0.60% | 0.3 | 0.30% | 0.15 |
25 | 0.90% | 0.225 | 0.60% | 0.15 | |||
5 | 2.00% | 0.1 | 2.00% | 0.1 | |||
ME-TopPette(1/10-100μl) | 10-100μl | 1μl | 100 | 0.80% | 0.8 | 0.15% | 0.15 |
50 | 1.00% | 0.5 | 0.40% | 0.2 | |||
10 | 3.00% | 0.3 | 1.50% | 0.15 | |||
ME-TopPette(1/20-200μl ) | 20-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | |||
20 | 3.00% | 0.6 | 1.00% | 0.2 | |||
ME-TopPette(1/50-200μl) | 50-200μl | 1μl | 200 | 0.60% | 1.2 | 0.15% | 0.3 |
100 | 0.80% | 0.8 | 0.30% | 0.3 | |||
50 | 1.00% | 0.5 | 0.40% | 0.2 | |||
ME-TopPette(1/100-1000μl) | 100-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | |||
100 | 2.00% | 2 | 0.70% | 0.7 | |||
ME-TopPette(1/200-1000μl) | 200-1000μl | 5μl | 1000 | 0.60% | 6 | 0.20% | 2 |
500 | 0.70% | 3.5 | 0.25% | 1.25 | |||
200 | 0.90% | 1.8 | 0.30% | 0.6 | |||
ME-TopPette(1/1000-5000μl) | 1000-5000μl | 50μl | 5000 | 0.50% | 25 | 0.15% | 7.5 |
2500 | 0.60% | 15 | 0.30% | 7.5 | |||
1000 | 0.70% | 7 | 0.30% | 3 | |||
ME-TopPette(1/2-10ml) | 2-10ml | 0.1ml | 10ml | 0.60% | 60 | 0.20% | 20 |
5ml | 1.20% | 60 | 0.30% | 15 | |||
2ml | 3.00% | 60 | 0.60% | 12 |
Eight chanel | Volume Range | Increment | Test Volume(μl) | Accuracy error | Precision error | ||
Model | % | μl | % | μl | |||
ME-TopPette(8/0.5-10μl) | 0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 |
5 | 2.50% | 0.125 | 2.50% | 0.125 | |||
1 | 4.00% | 0.04 | 4.00% | 0.04 | |||
ME-TopPette(8/ 5-50μl ) | 5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 |
25 | 1.50% | 0.375 | 1.00% | 0.25 | |||
5 | 3.00% | 0.15 | 2.00% | 0.1 | |||
ME-TopPette(8/50-300μl) | 50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 |
150 | 1.00% | 1.5 | 0.50% | 0.75 | |||
50 | 1.50% | 0.75 | 0.80% | 0.4 | |||
Twelve chanel | Volume Range | Increment | Test Volume(μl) | Accuracy error | Precision error | ||
Model | % | μl | % | μl | |||
ME-TopPette(12/0.5-10μl ) | 0.5-10μl | 0.1μl | 10 | 1.50% | 0.15 | 1.50% | 0.15 |
5 | 2.50% | 0.125 | 2.50% | 0.125 | |||
1 | 4.00% | 0.04 | 4.00% | 0.04 | |||
ME-TopPette(12/ 5-50μl) | 5-50μl | 0.5μl | 50 | 1.00% | 0.5 | 0.50% | 0.25 |
25 | 1.50% | 0.375 | 1.00% | 0.25 | |||
5 | 3.00% | 0.15 | 2.00% | 0.1 | |||
ME-TopPette(12/50-300μl) | 50-300μl | 5μl | 300 | 0.70% | 2.1 | 0.25% | 0.75 |
150 | 1.00% | 1.5 | 0.50% | 0.75 | |||
50 | 1.50% | 0.75 | 0.80% | 0.4 |